ISOLATING TRANSFORMERS FOR MEDICAL USE COMPLYING WITH CEI/IEC 61558-1/-2 /-2-15
Use: Power supply for medical use, designed to be connected permanently to the fixed wiring of the IT (impedant earth) power system.
Type: ISOLTUM (Three-Phase)
Power: 3kVA ≤ P ≤ 10kVA
Primary voltage: ≤ 1000V. Usual: 230V single-phase ; 400V three-phase.
Secondary voltage: ≤ 250V. Usual: 230V between phases.
Coupling (Three-phase): Star + Neutral // Star + Neutral.
Frequency: ≤ 500Hz. Usual: 50/60Hz.
Special characteristics: Impedance voltage ≤ 3 %. Inrush current ≤ 12In. (8 In as standard)
Winding: Class H, enamelled electrolytic copper, treated and polymerized in an oven, class H varnish, class H insulation, electrostatic shield between primary and secondary.
Connectivity: Inputs, outputs and shield on marked terminal blocks with insulators.